Surface Enhancement and Characterization of L-605 Cobalt Alloy Cardiovascular Stent by Novel Electrochemical Treatment

Prakash Sojitra, Ankur Raval, Devesh Kothwala, Arpit Jariwala, Haresh Kotadia, Subodh Adeshara, Girish Mehta

2009 · 13 citations · 5 references

Concepts

Abstract

Smooth surface is prerequisite for all implantable medical devices to increase biocompatibility. Electropolishing is an effective process for improving surface smoothness of the cardiovascular devices. In this study, electropolishing was employed on laser cut L605 Co-Cr alloy cardiovascular stents to increase surface smoothness. Acid descaling was used as a pretreatment before electropolishing for effective removal of metal oxides generated due to laser-cutting process. Effect of temperature during electropolishing on the rate of material removal was also evaluated. Passivation was carried out after electropolishing of stent to enhance its corrosion resistance in biological environment. The stent was characterized by Optical Microscopy for strut dimension analysis, Scanning Electron Microscopy and Atomic Force Microscopy for its surface smoothness, gravimetric analysis for evaluation of material removal uniformity, EDAX analysis for elemental composition and balloon expandability. It was found that elemental composition of stent remains unaltered after series of chemical treatments including the acid descaling and the electropolishing. Surface roughness of stent after electropolishing was reduced to significant level and such electropolished stents also demonstrated uniform expansion under specific required conditions. © Society for Biomaterials and Artificial Organs (India), 20081210-31.

References

5