Performance characteristics of the University of Texas TOFPET-I PET camera

Wei Wong, Nizar A. Mullani, E. A. Philippe, R. K. Hartz, D. Bristow, K. Yerian, J.M. Gaeta, N. Ketharnavaz

OSTI OAI (U.S. Department of Energy Office of Scientific and Technical Information) · 1986 · 19 citations · 0 references

Concepts

Abstract

The performance of the University of Texas TOFPET-I PET camera has been characterized. The camera has 720 fast cesium fluoride scintillators arranged in 5 rings to produce 9 slices images. The spatial resolution FWHM is found to be 9.0 mm, 11.0 mm and 12.0 at the center, 9 cm and 18 cm off center radius respectively. The time-of-flight resolution is about 9 cm. The true coincidence sensitivity for the outer and inner slices are about 8500 cps/ Ci/cc and 22,000 cps/ Ci/cc respectively for a 20 cm uniform source. The axial field-of-view spans 10.5 cm and allows the whole heart to be imaged at one time and a 3-D surface model of the heart to be displayed. The 3-D image is found to be more useful than the conventional slice image for interpretation and is being used for normal operation. Human studies have been carried out for both normal and MI subjects with YSRb. Fast 4 seconds dynamic images with 50% cardiac gating have been obtained.