Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · 2015 · 12 citations · 11 references
EngineeringBiomedical EngineeringBiomedical ImagingOpticsRemote SensingImagingOptical ImagingPolar Remote SensingPolarization OpticsPolarization-based ImagingBandwidthImaging PolarimetryMueller Matrix ImagingPolarimeter DesignPolarimetryPolarimetry InstrumentsPolarization ImagingSensor Resolution
Imaging polarimeters have been largely used for remote sensing tasks, and most imaging polarimeters are division of time or division of space Stokes polarimeters. Imaging Mueller matrix polarimeters have just begun to be constructed which can take data quickly enough to be useful. We have constructed a Mueller matrix (active) polarimeter utilizing a hybrid modulation approach (modulated in both time and space) based on a micropo- larizer array camera and rotating retarders. The hybrid approach allows for an increase in temporal bandwidth (instrument speed) at the expense of spatial bandwidth (sensor resolution). We present the hybrid approach and associated reconstruction schemes here. Additionally, we introduce the instrument design and some preliminary results and data from the instrument.
11
Polarized light. Fundamentals and applications
Edward Collett · Optical Engineering · 1993 · 368 citations
Liquid crystal polymer full-stokes division of focal plane polarimeter
Graham Myhre, Wei-Liang Hsu, Alba Peinado et al. · Optics Express · 2012 · 140 citations · Full text