Rapid Recovery of Mycobacteria from Clinical Specimens Using Automated Radiometric Technic

Choong H. Park, Deborah L. Hixon, Carolyn B. Ferguson, Sandra L. Hall, Cecilia C. Risheim, C. Barrie Cook

American Journal of Clinical Pathology · 1984 · 41 citations · 3 references

Concepts

Abstract

Automated radiometric technic (BACTEC Johnston Laboratories, Towson, MD) was compared with conventional mycobacterial culture procedure (Lowenstein-Jensen plus Gruft modification of Lowenstein-Jensen) in this study of 1,000 clinical specimens. In addition, 8-azaguanine inhibition was tested by radiometric technic as a rapid procedure for the differentiation of Mycobacterium tuberculosis from other mycobacterial species. A total of 59 mycobacteria was recovered. Of 28 clinically significant isolates (M. tuberculosis, M. kansasii, M. avium, M. fortuitum), the BACTEC system detected 26 (93%). Conventional methods recovered 23 (82%). The BACTEC system required an average of seven days to recover M. tuberculosis from smear-positive specimens compared with 18 days required by Lowenstein-Jensen or Gruft slants. From smear-negative specimens, the BACTEC detected M. tuberculosis in an average of 20 days versus 28 days by conventional procedure. All 20 isolates of M. tuberculosis were inhibited by 8-azaguanine, whereas 39 isolates of mycobacteria other than M. tuberculosis were not inhibited. The BACTEC system accomplishes more rapid recovery of mycobacteria and provides a higher yield than conventional methods.