PubMed · 2003 · 10 citations · 15 references
MedicineRadiation OncologyRadiation DosimetryScientific ComputingTreatment PlanningComputational ModelingMonte Carlo DosimetryMonte Carlo MethodsMonte Carlo SimulationSmall Field DosimetryTreatment Planning Dose CalculationDepth Dose DistributionMonte Carlo Dose ModelingMonte Carlo Dose SimulationSpatial Dose Distribution Analysis
PURPOSE: A Monte Carlo (MC) model of CyberKnife was developed as a quality assurance tool. The percentage depth dose (%dd) was verified by using this model. MATERIALS AND METHODS: An MC model was developed with Electron Gamma Shower version 4 (EGS4) in two steps: (1) a model of the CyberKnife treatment head and (2) a model of the collimator and phantom. The bremsstrahlung spectrum was calculated using the first model, and this spectrum was then used to calculate %dds with the second model. The calculated %dds for a large field (60 mm diameter) and three small fields (30, 15, and 5 mm diameter) were compared with those measured with a diamond detector. RESULTS AND DISCUSSION: The MC-calculated and measured %dd-curves for the 60 mm diameter field were in excellent agreement (<1.85%), thus confirming the validity of the model. Discrepancies between the calculated and measured %dd-curves increased with decreasing field size, with considerable discrepancy (11.62%) for the 5 mm diameter field due to lateral electron disequilibrium. Accurate dose can be determined with MC even in small fields. CONCLUSION: The MC technique can provide reliable standard data for accurate dose delivery with high-technology radiotherapies using small beams.
15
BEAM: A Monte Carlo code to simulate radiotherapy treatment units
D. W. O. Rogers, Bruce Faddegon, G Ding et al. · Medical Physics · 1995 · 1.4K citations
The Cyberknife: A Frameless Robotic System for Radiosurgery
John R. Adler, Steven D. Chang, Martin J. Murphy et al. · Stereotactic and Functional Neurosurgery · 1998 · 664 citations
Medicine, Neurosurgery, Surgery +15