Focal Osteolysis at the Junctions of a Modular Stainless-Steel Femoral Intramedullary Nail

D. M. Jones, J. Lawrence Marsh, James V. Nepola, Joshua J. Jacobs, Anastasia K. Skipor, Robert M. Urban, Jeremy L. Gilbert, Joseph A. Buckwalter

Journal of Bone and Joint Surgery · 2001 · 70 citations · 20 references

Abstract

The presence of corrosion products at the taper junctions suggests that particulate debris was a major factor in the etiology of the radiographic findings of osteolysis, periosteal reaction, and cortical thickening. Serum chromium levels were substantially elevated in the patients with a modular femoral nail, and such levels may serve as a marker of fretting corrosion of these devices.

References

20